Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative: h3, h6: investigation summary the product was returned to depuy synthes for evaluation.Visual inspection revealed one portion of the black radel handle broken from the angled acet insertr.Additionally, device exhibits signs of usage and the broken piece was not returned for evaluation.The failure of the radel handle either cracking or breaking is due to impaction forces when the surgeon strikes the handle anvil with a heavy object when positioning the shell.The force is transmitted through the anvil into the radel handle either via the anti-rotation pin or directly into the handle through the anvil.This damage can be consistent due to repeated use.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.Based on the evidence provided, it is reasonable to confirm the reported allegation, as a crack propagation can lead to the breakage observed.A dimensional inspection was not performed since it was not applicable to the complaint condition.A functional test was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the angled acet insertr would have contributed to the complained device issue.Based on the investigation findings, the potential cause is traced to an end of life problem, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|