| Catalog Number |
N-TSM375 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/28/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Patient reported "rupture", "migration of silicone gel", "visible axillary lymph nodes up to 15x5 mm in size".This record is for the left side.Device remains implanted.
|
| |
|
Event or Problem Description
|
|
Patient underwent re-endoprosthesis of the mammary glands.This record is for the left side.Device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional and/or corrected data: a5, a6, b5, b6, d6b, h6.Device evaluation:
visual analysis of the photographs identified:
rupture: observed, opening on the device but cannot perform an assessment of the opening as no microscopic analysis can be performed.Migration: unable to observe since it is not related to the device.No additional observations are performed.No further actions are required since no issue in the manufacturing of the device is observed.
|
| |
|