| Model Number |
4000 |
| Medical Device Problem Code |
Failure to Infuse (2340)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B3: unknown; no information has been provided to date.No information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filled as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
During the occlusion test, it was observed that instead of the plunger retracting during the occlusion phase, an audible ¿pop¿ was heard, after which the plunger stopped abruptly and displayed a ¿travel course error¿ message.The issue specifically occurred after the gauge filled up with the stopcock closed.There was no patient involvement and no harm was reported.
|
| |
|