Catalog Number 050-87215A
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/29/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
 
Event or Problem Description
It was reported that a peritoneal dialysis (pd) patient discovered a fluid leak on the inside of the cassette door of their cycler after ending their pd treatment.The patient reported receiving an air detected in cassette alarm during drain 1 of treatment and the treatment was cancelled.Fluid was discovered in the pump module and on the external of the cassette.The cause of the leak is unknown.The patient was advised to discontinue the use of their cycler to allow it to dry out until the next treatment.Upon follow up, it was confirmed that the patient is trained on performing stat drains, as well as manual peritoneal dialysis therapy if needed.There was no patient serious injury or medical intervention required as a result of this event.The patient is continuing peritoneal dialysis therapy with no further issues.The cycler set used by the patient was discarded and is not available for return for physical evaluation by the manufacturer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIBERTY CYCLER SET, SINGLE PATIENT CONN
Common Device Name
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
director, quality systems
900 w zaragosa drive suite d
pharr TX 78577
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key25364436
Report Number0008030665-2026-01103
Device Sequence Number16115605
Product Code FKX
UDI-Device Identifier00840861102709
UDI-Public00840861102709
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K173718
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number050-87215A
Device Lot Number25BR08115
Was Device Available for Evaluation? No
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/29/2026
Supplement Date Received by Manufacturer07/09/2026
Initial Report FDA Received Date06/02/2026
Supplement Report FDA Received Date07/10/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/21/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DELFLEX PD FLUID; LIBERTY SELECT CYCLER
Patient SexFemale
-
-