• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: K2M, INC. CHESAPEAKE Cervical-Titanium Ø3.5 mm Universal Drill, Screw Ø3.5x14 mm Screw; Orthopedic Manual Surgical Instrument

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

K2M, INC. CHESAPEAKE Cervical-Titanium Ø3.5 mm Universal Drill, Screw Ø3.5x14 mm Screw; Orthopedic Manual Surgical Instrument Back to Search Results
Catalog Number 3608-90080
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/06/2026
Type of Reportable Event Malfunction
Event or Problem Description
A company representative reported that the tip of a cascadia universal drill fractured intra-operatively.The procedure was completed successfully with no surgical delay and no adverse consequence to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHESAPEAKE Cervical-Titanium Ø3.5 mm Universal Drill, Screw Ø3.5x14 mm Screw
Common Device Name
Orthopedic Manual Surgical Instrument
Manufacturer (Section D)
K2M, INC.
600 hope pkwy
leesburg VA 20175
Manufacturer (Section G)
K2M, Inc.
600 hope pkwy se
leesburg VA 20175
Manufacturer Contact
catherine devarenne
2 pearl court
allendale, NJ 07401
+1(201)760-8000
MDR Report Key25365381
Report Number3004774118-2026-00146
Device Sequence Number10679746
Product Code LXH
UDI-Device Identifier10888857083011
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number3608-90080
Device Lot NumberCPMF
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer05/11/2026
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/06/2026
Supplement Date Received by Manufacturer08/07/2026
Initial Report FDA Received Date06/02/2026
Supplement Report FDA Received Date09/02/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/10/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-