|
Evaluation of the returned velocity delivery microcatheter (velocity) confirmed that the catheter was fractured.Due to lack of signs of torquing or stretching at the fractured location, this damage was likely a kink that worsened to a fracture.If the velocity is retracted against resistance at an angle, damage such as a kink and subsequent fracture may occur.Based on the reported event, the root cause of resistance could not be determined.Further evaluation revealed additional fractures along the catheter shaft.This damage was incidental to the reported complaint, and the root cause could not be determined.Penumbra devices are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|