Catalog Number 3505590BC
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 02/02/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient underwent a breast augmentation revision surgery with implantation of a 590cc mentor memorygel breast implant prosthesis.Post-operatively, the patient reported experiencing a hard right breast with some changes.She desired a size reduction and an implant plane change from submuscular to subglandular as she had animation issues with the breasts.An ultrasound was conducted and confirmed a ruptured right implant.A consultation with a medical professional was conducted.As a result, the patient underwent bilateral exchange with motiva (non-mentor) implants on (b)(6) 2026.This report is for the left breast prosthesis.
 
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: animation deformity.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
henry boden
3041 skyway circle north
irving, TX 75038
9497882078
MDR Report Key25368175
Report Number1645337-2026-06358
Device Sequence Number9847661
Product Code FTR
UDI-Device Identifier00081317006992
UDI-Public(01)00081317006992(17)220512(10)7467945
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/12/2022
Device Catalogue Number3505590BC
Device Lot Number7467945
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/13/2026
Initial Report FDA Received Date06/02/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/15/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age29 YR
Patient SexFemale
Patient EthnicityHispanic
-
-