• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET GMBH OTESUS OPERATING TABLE SYSTEM; TABLE, OPERATING-ROOM, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET GMBH OTESUS OPERATING TABLE SYSTEM; TABLE, OPERATING-ROOM, AC-POWERED Back to Search Results
Model Number 116001D0
Medical Device Problem Code Device Tipped Over (2589)
Health Effect - Clinical Code Crushing Injury (1797)
Date of Event 05/13/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional information will be provided following the conclusion of the investigation.D10 concomitant products: 115030d0 - modular universal tabletop s/n (b)(6).E1 initial reporter: (b)(6).
 
Event or Problem Description
Getinge became aware of an issue with one of our table 116001d0 - or table column, manoeuvrable.As it was stated, while using the down function in the reverse trendelenburg position, the column rose several centimeters above the floor before suddenly dropping back down, striking the surgeon¿s foot.The injury sustained required surgical intervention.No other injuries were reported.We have decided to report the issue due to the serious injury sustained by the surgeon.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OTESUS OPERATING TABLE SYSTEM
Common Device Name
TABLE, OPERATING-ROOM, AC-POWERED
Manufacturer (Section D)
MAQUET GMBH
kehler strasse 31
rastatt
GM 
Manufacturer (Section G)
MAQUET GMBH
kehler strasse 31
rastatt
GM  
Manufacturer Contact
diana gutekunst
kehler strasse 31
rastatt 
GM  
MDR Report Key25372449
Report Number8010652-2026-0000041
Device Sequence Number21485168
Product Code FQO
UDI-Device Identifier04046768122161
UDI-Public(01)04046768122161(11)221129
Combination Product (Y/N)N
Initial Reporter CountryIS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number116001D0
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/13/2026
Initial Report FDA Received Date06/03/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/29/2022
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
115030D0
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-