| Model Number |
DDU-C2300 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer reported that their aed's asi is blank, and the aed will not power on.The customer did not report this occurring during patient use.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis of the aed's log files identified that the customer's failure to promptly address red asi warnings reduced battery life expectancy.On (b)(6) 2025, the aed identified that the aed was placed into service with out the pads attached and attempted to alert the user by flashing its red asi and chirping.The aed continued to flash and chirp until on (b)(6) 2025 when the battery pack became completely depleted.There was no evidence in the electronic logs of any human interaction to address the aeds condition until on (b)(6) 2026 when a replacement battery pack was inserted into the aed.The review identified that the aed functioned as designed and can stay in service.
|
| |
|