| Model Number |
12673-03 |
| Medical Device Problem Code |
Malposition of Device (2616)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/14/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot specific issue.Based on the information received, the investigation determined that the reported issue appears to be related to operational context.In this case, it is likely that interaction with the patient anatomy or friable femoral vessel contributed to the reported difficulty.Based on the results of the complaint investigation, there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that this was a venotomy closure of a right common femoral vein using a proglide device after an interventional ablation procedure with an 8f sheath.Reportedly, when removing the plunger, the suture continued to come out of the device as there was no tension felt when pulling back.The suture was manually cut and an attempt was made to harvest the suture; however, the knot was unraveled.A new proglide device was used to achieve hemostasis.There were no adverse patient effects and no reported clinically significant delay in the procedure or therapy.No additional information was provided.
|
| |
|