W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
|
Back to Search Results |
|
| Catalog Number |
CXT281412 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Aneurysm expansion (4597); Clinical Endoleak (4843)
|
| Date of Event |
03/06/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
On (b)(6) 2025, this patient underwent endovascular treatment of an aortoiliac aneurysm and a right common iliac artery aneurysm (rcia)was implanted with gore® excluder® conformable aaa endoprosthesis devices and gore® excluder® iliac branch endoprosthesis devices.Pre-procedure computed tomography angiography (cta) performed on (b)(6) 2025, determined the maximum diameter of the abdominal aorta measured 56mm with a maximum diameter proximal landing zone of 25.4mm.The maximum diameter of the rica and lcia were not provided.Successful access, delivery, and accurate deployment of the devices to their intended location, and retrieval of the delivery system and technical success were reported.There were no corrective procedure(s) or complications related to the device or procedure.The patient tolerated the procedure and was discharged to home (self-care) on (b)(6) 2025.On (b)(6) 2025, the patient underwent follow-up cta and was reported to have a type ii endoleak involving the inferior mesenteric artery (ima) and lumbar arteries.There is no loss of patency to a gore® device and no device integrity events reported.This event is noted to be on going and no treatment was provided.On (b)(6) 2025, the patient underwent follow-up color duplex ultrasound determined the type ii endoleak persisted.Additionally, the maximum diameter of the abdominal aorta measured 23.5mm, the rcia diameter measured 9.1mm and the lcia diameter measured 11.3mm.There is no loss of patency to a gore® device and no device integrity events reported.This event is noted to be on going and no treatment was provided.On (b)(6) 2026, the patient underwent reintervention and coil embolization was performed.The patient received onyx liquid embolic system injection into the lumbar arteries and median sacral artery second.Additionally, multiple large detachable packing coils were deployed within the aneurysm sac".Lastly, they had a coil embolization with embold 6mm coils into the inferior mesenteric artery.The patient tolerated the procedure.On (b)(6) 2026, the patient underwent follow-up cta which determined the type ii endoleak persisted, the maximum diameter of the abdominal aorta measured 62mm, no diameters were provided for the rcia or lcia.There is no loss of patency to a gore® device and no device integrity events reported, the event was due to disease progression.This event is noted to be on going and no treatment was provided at this time but would be required at a later date.On (b)(6) 2026, the patient underwent reintervention this date.At the end of the procedure, a complete angiography was performed where it stated the following, "repeat angiography performed through the inferior mesenteric artery demonstrated complete occlusion of the endoleak with no residual filling of the aneurysm sac or identified outflow vessels.It was reported that the type ii endoleak had resolved.The patient tolerated the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Search Alerts/Recalls
|
|
|