• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY EVO/EVO+ VISIAN IMPLANTABLE COLLAMER LENS; PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY EVO/EVO+ VISIAN IMPLANTABLE COLLAMER LENS; PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number VICM5 13.2 (-7.00)
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Inflammation (1932)
Date of Event 04/14/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H11-manufacturer narrative: toxic anterior segment syndrome (tass) is an acute sterile postoperative inflammation that can occur after uncomplicated or complicated intraocular surgery.While improper sterilization and contamination of surgical instruments remains the most common risk factor associated with tass, ocular viscoelastic, and improper preparation of antibiotics for intracameral injection may also lead to tass.An exact cause could not be determined as the event could be multifactorial in nature including patient and/or procedure related factors.Claim # (b)(4).
 
Event or Problem Description
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced mild toxic anterior segment syndrome.Treatment was performed and problem resolved.Lens remains implanted.Cause of the event was reported as other: tass.No further information has been provided.If additional information is received, a supplemental medwatch report will be submitted.
 
Additional Manufacturer Narrative
Additional information: b5 - describe event or problem - a facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced mild toxic anterior segment syndrome.No diagnostic culture was performed.Treatment and surgical management was performed.Problem was resolved.Lens remains implanted.Cause of the event was reported as other: tass.No further information has been provided.If additional information is received, a supplemental medwatch report will be submitted.D4 - serial #: (b)(6), expiration date: 12/31/2028, primary unique device identifier (udis) #: (b)(4).H4 - device manufacturer date: 01/07/2026.H6: investigation type 4110: a lens work order search was performed.No additional similar complaint type events within associated lots were found.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVO/EVO+ VISIAN IMPLANTABLE COLLAMER LENS
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene muniz
1911 walker avenue
monrovia, CA 91016
MDR Report Key25379536
Report Number2023826-2026-00931
Device Sequence Number9478989
Product Code MTA
UDI-Device Identifier00840311300877
UDI-Public(01)00840311300877(11)260107(17)281231
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberVICM5 13.2 (-7.00)
Device Catalogue NumberVICM5 13.2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/27/2026
Supplement Date Received by Manufacturer06/07/2026
Initial Report FDA Received Date06/03/2026
Supplement Report FDA Received Date06/10/2026
Date Device Manufactured01/07/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL: UNK, LOT# UNK; FOAM TIP PLUNGER: UNK, LOT #: UNK; INJECTOR MODEL: UNK, LOT#: UNK
Outcome Attributed to Adverse Event Required Intervention;
Patient Age47 YR
Patient SexFemale
-
-