|
The device was not returned for analysis; therefore, a technical analysis could not be performed.Good faith effort attempts were made to try and retrieve the device, however response was not received.The manufacturing process for this electrode model included extensive inspections to ensure that the finished product met specifications prior to distribution, including validation of sterility and a series of visual, functional, and electrical tests.There is no indication that the referenced event was related to the manufacturing process.A risk review was completed and confirmed that the event of infection was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Review of labeling determined that the complaint situation was listed in the manual.There was no indication in the complaint that the product was not used in accordance to labeling.The manual was unlikely to be the cause of the reported complaint; translation, wording, or graphics does not require further review.A similar complaint review was completed using similar complaint review tool, which reviewed ar-p code 9151: infection and did not identify any emerging trends or similar complaints that require escalation.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
|