• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. EQUINOXE; PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. EQUINOXE; PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Catalog Number 320-10-00
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Pain (1994); Joint Dislocation (2374); Aseptic loosening (4883)
Date of Event 02/09/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a 71 yo male patient, who had a left rtsa, underwent a revision procedure.The reported information indicated the patient had a successful shoulder replacement, however the stem has since loosened resulting in a dislocation.Additional information from the clinical study notes the patient's dog jumpled on him and he felt a sudden onset of pain.X-ray images showed dislocation and movement of the humeral component.As such, a single stage revision was undertaken, so that the stem could be cemented.They were revised to a rtsa, with a cemented stem for rotational stability and an increase in liner size to 38+2.5mm for increased lengthening.The implants were removed and replaced with no issues.There were no device breakages or surgical delays during the procedure.The patient was last known to be in stable condition following the event.No x-rays were able to be obtained.The explanted devices are available for return.No further information.This is 1 of 4 reports.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE
Common Device Name
PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 ct
gainseville FL 32653
Manufacturer Contact
julia goldberg
2320 nw 66 court
gainseville, FL 32653
3523771140
MDR Report Key25382179
Report Number1038671-2026-00710
Device Sequence Number9204599
Product Code KWT
UDI-Device Identifier10885862086419
UDI-Public10885862086419
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K063569
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number320-10-00
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/11/2026
Initial Report FDA Received Date06/03/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/07/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age71 YR
Patient SexMale
-
-