• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALTATEC ALTATEC; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALTATEC ALTATEC; DENTAL IMPLANT Back to Search Results
Model Number C1086.4311
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 02/18/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Biohorizons issue: record failed with reason of duplicate number.Honeywell sparta vendor response: more than one registered site using the registration number 1060818 was assigned the same number.Therefore, causing the records to have a duplicate.Various sites having same value in "assigned number" for two or more registered sites will create the duplicate report number using between the emdr with different registered sites, when on start of the year, last used sequence number will reset to 1 for all the registered site.Corrective action by biohorizons: updated the last used sequence number field value to 60000 for rs-0002 and with 80000 for rs-0003, as rs-0000, rs0002 and rs0003 share the same registration number.The vendor suggested as a workaround by creating a starting sequence range high enough to prevent duplication.For importer files, the sequence number begins at 60000 for 2026.Clear the duplicate mfr report number and resubmit the record to capture the new last sequence number beginning with 60nnnn.The clearance would be either through automation or manually.Preventative action: work with vendor honeywell/sparta to create an automation to update the last used sequence number in the future and prevent duplication records.
 
Event or Problem Description
Implant failed to osseointegrate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTATEC
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ALTATEC
maybachstrasse 5
wimsheim 71299
GM  71299
MDR Report Key25382507
Report Number1060818-2026-60398
Device Sequence Number9324643
Product Code DZE
UDI-Device IdentifierE219C108643111
UDI-PublicE219C108643111
Combination Product (Y/N)N
Initial Reporter CountryUK
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 03/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberC1086.4311
Device Catalogue NumberC1086.4311
Device Lot NumberUNKNOWN
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date02/18/2026
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date06/03/2026
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-