| Model Number |
AFR-00001 |
| Medical Device Problem Codes |
Use of Device Problem (1670); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Muscle Weakness/Atrophy (1967)
|
| Date of Event |
05/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during a cardiac ablation procedure, the sphere 9 catheter malfunctioned and was removed.Upon direct observation there was a solid white substance observed inside the catheter tip.The substance was confirmed by pathology to be fibrin.It was further reported that the patient experienced transient arm weakness and was noted that it may have been a tia (transient ischemic attack), however computed tomography angiography (cta) was negative and there were no residual neurological issues.The procedure was not able to be completed.No further patient complications have been reported as a result of this event.
|
| |
|