H3:product analysis:evaluation determined that the distal bearing is detached.The likely cause of the failure is the worn of the input shaft and output shaft.It was also noted that the spacer on the tube is bent, the color band is faded or illegible, and the laser marks are faded or illegible.G3:aware date used as date of analysis performed date as the event is reported based on evaluation findings.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|