Catalog Number 010000589
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 05/20/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10: comp rvs cntrl 6.5x30mm st/rst cat#:115396 lot#: 67418074.G2: foreign - event occurred in japan.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event or Problem Description
It was reported that the glenosphere head became dislodged following a shoulder arthroplasty, necessitating reoperation approximately three months post-implantation.During postoperative follow-up, the dislodgement was identified, and during the reoperation the central screw was retightened and the glenosphere head was reinserted.The reported cause was that the central screw had not been inserted to a sufficient depth during the initial surgery.The patient underwent reoperation with successful completion of the procedure.Attempts have been made and no further information has been provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMP RVRS 25MM BSPLT HA+ADPTR
Common Device Name
SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key25391150
Report Number0001825034-2026-01594
Device Sequence Number10718682
Product Code PHX
UDI-Device Identifier00880304532465
UDI-Public(01)00880304532465(17)350813(10)67243846
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K193373
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/04/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number010000589
Device Lot Number67243846
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/20/2026
Initial Report FDA Received Date06/04/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/13/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-