| Model Number |
CYF-VHR |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.Based on the results of the investigation, a definitive root cause of the detached device bending section could not be determined, however, the issue was likely the result of component failure due to repetitive use stress, handling/excessive applied force and or external factors.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
The cysto-nephro videoscope was returned to the service center with a report of a defective controller.This does not indicate an mdr reportable event, however, a reportable failure was found during testing at the service center.During inspection of the device, a detached bending section was found, likely due to stress.There was no patient involvement, and no harm was reported as a result of the event.
|
| |
|