| Model Number |
CI-1601-05 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
06/04/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The recipient is reportedly experiencing poor performance.A ct scan confirmed electrode migration.Revision surgery is scheduled.
|
| |
|
Additional Manufacturer Narrative
|
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: section b.3, d.6b, d.9.Corrected information: section b.1, b.2 & h.1.Advanced bionics considers the investigation into this reportable event as closed.The recipient's device was explanted.The recipient was re-implanted with another advanced bionics cochlear device.The external visual inspection revealed the electrode was severed and silicone damage was observed on the top and bottom cover of the device.These are believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The condition of the electrode prevented an electrical test from being performed.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
| |
|