H3: product analysis: evaluation confirmed that burr could not be locked into the attachment.The likely cause was identified as due to detached distal bearing.It was also noted that two bearings spacers were worn, there was excessive corrosion inside the attachment which caused pitting in the drive shafts and tactile sleeve, the laser markings were faded and the dial was frozen.B3: date of notification: (b)(6) 2026 (date of event or estimated date of incident not provided to manufacturer).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|