| Model Number |
RLM314K3 |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A steris service technician arrived onsite following the reported event to inspect the unit and found that the avc-x of the rack was not operating properly.To resolve the issue, the technician reset the avc-x.The unit was tested, confirmed to be operating according to specification, and returned to service.No additional issues have been reported.
|
| |
|
Event or Problem Description
|
|
The user facility reported that prior to a patient procedure the monitor to their vividimage 4k surgical display was showing lines.No report of injury.
|
| |
|