ARROW INTERNATIONAL LLC ARROW PSI KIT: 9 FR; INTRODUCER CATHETER Back to Search Results
Catalog Number AK-09903-J
Medical Device Problem Codes Crack (1135); Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/08/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Continuation of d11: r preparation for cardiopulmonary bypass.(b)(4).
 
Event or Problem Description
It was reported that "during the procedure of gaining access to the patient's right internal jugular (rij) vein with the introducer needle and 5ml syringe, the anesthesiologist became aware there was a blood and air mixture in the syringe.The anesthesiologist confirmed with ultrasound that the needle tip was appropriately in the middle of the vessel and still having a blood/air mixture.Upon closer review of the needle, it was noted there was a crack in the plastic hub portion of the steel introducer needle.The anesthesia team immediately proceeded to use a new 9fr introducer sheath without any additional issue.The patient's rij had to be punctured again.There were no reports of patient injury, harm, or adverse health consequences associated with this event.The patient's condition was reported as "fine".No additional medical intervention was reported beyond removal of the affected device, and replacement with another device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW PSI KIT: 9 FR
Common Device Name
INTRODUCER CATHETER
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL DE MEXICO S.A. DE C.V.
ave. washington 3701
colonia panamericana, chihuahua
chihuahua 31200
MX   31200
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key25411608
Report Number9680794-2026-00427
Device Sequence Number5840328
Product Code OFC
UDI-Device Identifier40801902001274
UDI-Public(01)40801902001274(11)260108(17)300731(10)33F25J0830(21)
Combination Product (Y/N)N
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberAK-09903-J
Device Lot Number33F25J0830
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/20/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/11/2026
Supplement Date Received by Manufacturer06/09/2026
Initial Report FDA Received Date06/05/2026
Supplement Report FDA Received Date06/18/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/08/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PATIENT TO RECEIVE AN UNKNOWN AMOUNT OF HEPARIN FO