Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative: h3, h6: investigation summary the product was returned to depuy synthes for evaluation.Visual inspection revealed the overall surface of the baseplate inserter rod with signs of usage and its threaded tip was found stripped.The observed condition of the threaded tip was consistent with off-axis assembly and impacting the device before it was fully seated.Properly handling and attention to the approved use of the device diminishes the risk of failure.As per inhance¿ shoulder system surgical technique, the next precautions need to be followed: 1) "thread the baseplate inserter rod into the central hole of the baseplate until the handle of the baseplate inserter rod is snug with the baseplate inserter handle.Care should be taken not to over-tighten"; and 2) "place the baseplate into the prepared glenoid over the glenoid pin, taking care to orient the peripheral holes where desired to achieve the best fixation.When the desired orientation is achieved, impact the baseplate into place.After the baseplate is fully seated, use the baseplate inserter handle to provide counter-torque and unthread the baseplate inserter rod".A dimensional inspection was unable to be performed due to the post-manufacturing damage.A functional test was unable to be performed due to the post-manufacturing damage.The overall complaint was confirmed as the observed condition of the baseplate inserter rod would have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|