| Model Number |
WA2T412A |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/15/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: distal end lens displacement due to a component failure of the distal end lens.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
The customer returned the rigid endoscope telescope to the service center for a separate issue.During the device analysis, the investigation indicates that a distal end lens displacement occurred.There was no patient involvement.
|
| |
|
Search Alerts/Recalls
|