OLYMPUS WINTER & IBE GMBH VIDEO TELESCOPE "ENDOEYE", 5 MM, 30°, NTSC, AUTOCLAVABLE, SHORT; RIGID VIDEO LAPAROSCOPE
|
Back to Search Results |
|
| Model Number |
WA50023S |
| Medical Device Problem Code |
Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/14/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, the most probable cause of the malfunction was traced to chipped outer tube.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
The customer returned the rigid video laparoscope to the service center for evaluation related to a separate issue.During device evaluation, the repair center technician identified the device had twisted image.There was no patient involvement.
|
| |
|