Per the instructions for use (ifu), structural valve deterioration (wear, fracture, calcification, leaflet tear/tearing from the stent posts, leaflet retraction, suture line disruption of components of a prosthetic valve, thickening, stenosis), and device degeneration are known potential risks associated with the use of bioprosthetic heart valves.The ifu cautions that accelerated deterioration of the valve may occur in patients with an altered calcium metabolism.Long-term durability has not been established for the valve.Regular medical follow-up is advised to evaluate valve performance.Structural valve deterioration (svd) may be manifested as stenosis with thickened leaflets.Svd may be mild and not require any intervention or it may be moderate to severe.It can cause the heart to work harder to eject blood from the ventricle.Depending on the severity it could be an indication for valve replacement or medical intervention.Tissue calcification is a common failure mode.The mechanisms for bioprosthetic heart valve tissue calcification are not yet fully understood.Many factors can contribute to the onset and propagation of calcification including patient-related (e.G., patient age, disease state, immune status, and other co-morbidities), pharmacological, and intrinsic properties of the valve itself.It is widely understood that patients with chronic renal disease and prior history of calcific stenosis of the native valve may be predisposed to bioprosthetic calcification.During the manufacturing process, all sapien thv valves are 100% visually inspected for defects and 100% functionally tested for proper coaptation under physiological backpressure conditions prior to release for distribution.Therefore, it is highly unlikely that a manufacturing defect or device malfunction contributed to the event.In this case, there was no allegation or indication a manufacturing non-conformance contributed to this product problem.The exact cause of the event was unable to be determined but may have been related to progression of disease, implant duration, and patient factors such as a history of hypertension, hyperlipidemia, aortic stenosis and/or the mechanisms described above.A review of edwards lifesciences risk management documentation was performed for this case.The reported event is an anticipated risk of the transcatheter heart valve procedure, additional assessment of this adverse event is not required at this time.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.As such, neither a product risk assessment, nor corrective or preventative actions are required at this time.
|
|
As reported by the partner 3 valve in valve (p3a) clinical study, a patient underwent a transfemoral tavr procedure with a 23mm sapien 3 valve inside a pre-existing edwards surgical perimount valve in aortic position.Approximately 7 years and 10 months later, the patient was admitted to the hospital with severe worsening of shortness of breath.An echo showed ava 0.3, mean gradient 97mmhg, and peak velocity of 621cm/sec.There was note of severe central aortic regurgitation as well.The patient underwent explant of the valve with savr.According to the medical record review, the patient had experienced worsening dyspnea and bilateral lower extremity edema for several months.The symptoms worsened 4 days prior to admission, echocardiogram revealed severe aortic stenosis and reduced ejection fraction.The patient developed acute hypoxic respiratory failure, pulmonary edema and hematuria.The patient was managed with diuretics, steroids, and inhalation therapy.Cardiology recommended urgent evaluation for redo savr.One week post admission the patient underwent redo aortic valve replacement and was discharged 6 days later.
|