• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ECHELON ENDOPATH; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ECHELON ENDOPATH; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number GST45W
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Ecchymosis (1818); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 05/18/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Date sent 6/8/2026.D4 batch # unk.Investigation summary: an analysis of the product could not be performed since a physical sample was not received for evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this lot.Additional information: dr.Performed a gastric bypass surgery where everything went well to the naked eye, when he is going to perform the postoperative control the same day in the afternoon note to the patient with his altered hemodynamics parameters, they perform an endoscopy to check if there is any problem with the surgery and it is noticed that close to the line of staples of the gastric pouch and the gastrojejunal anastomosis that he performs manually the patient presents hematomas and clots, therefore i require that you please check the batches of the refills used in this patient to see if there is any problem with them or if this batch has been reported previously and see if our refills may have failed.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: how was the post-op bleeding identified? how many days postoperative was the bleeding identified? what was the total estimated blood loss (ml)? was a transfusion required? how was the bleeding addressed? what was the pre and postoperative anti-coagulant and pain management protocol? was the bleeding intraluminal or extraluminal? were the staples the appropriate b-shape form? any clips, clamps, or graspers near the area where the device was being fired? did the healthcare professional wait 15 seconds after closing the device before firing? were there any issues with opening the device? if yes, what trouble shooting steps were used to open the device? were there any issues noted with staple formation at any point throughout the care of the patient? if so, please describe the shape and location.What color reloads were used and what was the sequence of the firings? was a leak test performed? if so, what type and what was the result? please provide a timeline of events.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during a gastric bypass the doctor said everything went well at first glance, when he is going to perform the postoperative control the same day in the afternoon he notices the patient with her hemodynamic parameters altered, they perform an endoscopy to check if there is any problem with the surgery and he realizes that close to the staple line of the gastric pouch and the gastrojejunal anastomosis that he performs manually, the patient has bruises and clots.Was there any harm to the reported patient or user? yes.Did the patient/user require any medical or surgical intervention as a result of the event? yes.Description: the patient had to undergo surgery again, since in the postoperative period the doctor performed an endoscopy and showed that the patient had bruises and clots in the gastrojejunal anastomosis, so it was decided to operate on her again.Did the patient/user require prolonged hospitalization as a result of the event? unknown.What is the patient/user status after the event? now it's okay.Was there a significant delay?.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHELON ENDOPATH
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
*  
3035526892
MDR Report Key25428825
Report Number3005075853-2026-04334
Device Sequence Number10310466
Product Code GDW
UDI-Device Identifier10705036014928
UDI-Public(01)10705036014928(17)280731(10)724D66
Combination Product (Y/N)N
Initial Reporter CountryCI
PMA/510(K) Number
K163454
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberGST45W
Device Lot Number724D66
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/19/2026
Initial Report FDA Received Date06/08/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/22/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-