| Catalog Number |
15-371 |
| Medical Device Problem Codes |
Device Appears to Trigger Rejection (1524); Material Rupture (1546)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported "capsular contracture baker grade iii".Later healthcare professional reported "rupture" discovered at explant.This record is for the left side.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Continued e1 - alternate email address: (b)(6).A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been/will be requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iii; (rupture discovered at explant).
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "capsular contracture baker grade iii".Later healthcare professional reported "rupture" discovered at explant.This record is for the left side.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b6, d9, h3, h6.Device evaluation:
per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required:
rupture: observed broken device through microscopic inspection assessed as fold flaw opening and observed a missing piece of shell through microscopic inspection assessed as inconclusive.No further actions are required as it is not related to the manufacturing process.Capsular contracture: unable to observe through visual inspection as it is a physiological phenomenon.None of the other observations performed during the device analysis (non-penetrating nick) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|
| |
|