• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) Style 15 Silicone Gel Filled Breast Implant; prosthesis, breast, noninflatable, internal, silicone gel-filled

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) Style 15 Silicone Gel Filled Breast Implant; prosthesis, breast, noninflatable, internal, silicone gel-filled Back to Search Results
Catalog Number 15-371
Medical Device Problem Codes Device Appears to Trigger Rejection (1524); Material Rupture (1546)
Health Effect - Clinical Code Capsular Contracture (1761)
Type of Reportable Event Serious Injury
Event or Problem Description
Healthcare professional reported "capsular contracture baker grade iii".Later healthcare professional reported "rupture" discovered at explant.This record is for the left side.Device has been explanted.
 
Additional Manufacturer Narrative
Continued e1 - alternate email address: (b)(6).A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been/will be requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iii; (rupture discovered at explant).
 
Event or Problem Description
Healthcare professional reported "capsular contracture baker grade iii".Later healthcare professional reported "rupture" discovered at explant.This record is for the left side.Device has been explanted.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: b6, d9, h3, h6.Device evaluation: per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: rupture: observed broken device through microscopic inspection assessed as fold flaw opening and observed a missing piece of shell through microscopic inspection assessed as inconclusive.No further actions are required as it is not related to the manufacturing process.Capsular contracture: unable to observe through visual inspection as it is a physiological phenomenon.None of the other observations performed during the device analysis (non-penetrating nick) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Style 15 Silicone Gel Filled Breast Implant
Common Device Name
prosthesis, breast, noninflatable, internal, silicone gel-filled
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS 
Manufacturer (Section G)
Allergan (Costa Rica)
900 parkway global park
zona franca
la aurora de heredia
CS  
Manufacturer Contact
terry ingram
2525 dupont drive
irvine, CA 92612
+1(847)936-6324
MDR Report Key25429045
Report Number9617229-2026-10746
Device Sequence Number10310025
Product Code FTR
UDI-Device Identifier10888628001473
UDI-Public(01)10888628001473(17)170811
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/11/2017
Device Catalogue Number15-371
Device Lot Number2343229
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/20/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/18/2026
Supplement Date Received by Manufacturer08/20/2026
Initial Report FDA Received Date06/08/2026
Supplement Report FDA Received Date09/08/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/11/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient Weight183 LB
Patient RaceWhite
-
-