CHRISTOPH MIETHKE GMBH & CO KG M.BLUE PLUS VALVE; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FX804T
Medical Device Problem Codes Obstruction of Flow (2423); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 05/04/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Visual inspection during the investigation, scratches on the outer housing of the valves were determined.Permeability test a permeability test has shown that both valves are permeable.Computer controlled test to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical position.The results show that the valves are operating within the specified tolerances in their respective relevant position.Adjustment test the valves were tested and both valves are adjustable in all pressure settings.Braking force and brake function test the brake functionality test has shown that the brake function operates as expected and the braking force is within the given tolerance.Internal inspection after dismantling of the valves, no deposits were found.Results in advance, we would like to point out that the product sent in was not inserted in liquid at the time of delivery.Dried organic deposits can influence the function of the products, which can affect the results.Despite this, we have examined the valve.Based on our investigation results, we cannot determine functional deviations or other abnormalities in the product.How the abovementioned functional impairment occurred is not clear to us at the time of examination.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Event or Problem Description
It was reported that an m.Blue plus(#fx804t) was implanted on (b)(6) 2023.According to the complainant, the shunt system caused an under-drainage and had adjustment difficulties.The patient underwent a revision procedure performed on (b)(6) 2026.No patient complications were reported as a result of the revision procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M.BLUE PLUS VALVE
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key25430452
Report Number3004721439-2026-00151
Device Sequence Number10310075
Product Code JXG
UDI-Device Identifier04041906504793
UDI-Public(01)04041906504793(11)240718(17)290716(10)20076582
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K192266
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFX804T
Device Catalogue NumberFX804T
Device Lot Number20076582
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/13/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/13/2026
Initial Report FDA Received Date06/08/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/18/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age33 YR
Patient SexFemale
Patient Weight90 KG
-
-