• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL MITRACLIP G5 DELIVERY SYSTEM; MITRAL VALVE REPAIR DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL MITRACLIP G5 DELIVERY SYSTEM; MITRAL VALVE REPAIR DEVICES Back to Search Results
Model Number CDS0801-NTW
Medical Device Problem Codes Premature Activation (1484); Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/14/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.Based on available information, the cause of the reported unable to open clip and premature activation were unable to be determined.The reported unexpected medical intervention was a result of case-specific circumstances.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Event or Problem Description
It was reported on (b)(6) 2026, that the patient presented with grade 4+ functional mitral regurgitation (mr) and restricted posterior leaflet for a mitraclip procedure.During the procedure, the steerable guide catheter (sgc) was advanced into the left atrium, followed by advancement of the clip.After advancement, an attempt was made to open the clip; however, the clip failed to open despite troubleshooting maneuvers.As a result, an attempt was made to remove the clip from the anatomy.During this process, it was noted that the clip had detached from the clip delivery system (cds).A snare was utilized to retrieve the clip, followed by balloon dilation of the septum to facilitate withdrawal of the snared clip into a non-abbott sheath.The clip was successfully removed from the anatomy.During preparation of second clip, the clip was unable to pass establish final arm angle test (efaa) as the clip jumped open to 120 degree.The clip was replaced with another device to complete the procedure.The mr was reduced to grade <1 post procedure.There was no clinically significant delay reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G5 DELIVERY SYSTEM
Common Device Name
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key25430968
Report Number2135147-2026-03854
Device Sequence Number13309241
Product Code NKM
UDI-Device Identifier05415067050748
UDI-Public(01)05415067050748(17)270115(10)50718R1020
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P100009
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberCDS0801-NTW
Device Catalogue NumberCDS0801-NTW
Device Lot Number50718R1020
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/14/2026
Initial Report FDA Received Date06/08/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/18/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
STEERABLE GUIDE CATHETER
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-