| Model Number |
9-ITV07F180/80 |
| Medical Device Problem Codes |
Difficult to Insert (1316); Retraction Problem (1536)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/11/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
It is unknown if the device is returning for analysis.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, a 8-6mm amplatzer pda occluder was chosen for implant using an unknown delivery system.During the procedure, it was reported that the device was not able to be retrieved back to the sheath.No additional information provided.
|
| |
|
Event or Problem Description
|
|
It was reported that on 11 may 2025, an 10-8mm amplatzer pda occluder was chosen for implant using a 7f amplatzer torqvue delivery system.It was reported that the device could not be retracted back into the sheath during an attempt to retrieve it for size exchange.No patient complications or adverse events were reported during or after the procedure, and the patient experienced no clinical signs or symptoms.The patient remained hemodynamically stable throughout the procedure, with no clinically significant delay reported, and was stable and discharged post-procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of a loader to sheath handoff failure and recapture the occluder device was reported.A returned device assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all defined manufacturing specifications.Additionally, the lot was reviewed for similar complaints, and there is no indication of a lot-specific product issue.Based on the available information, the cause for the reported loader to sheath handoff failure could not be determined.It is possible that procedural circumstances (insufficient tightening) contributed to the event, however this cannot be determined.There is no indication of a product quality issue with respect to labeling, design, or manufacturing of the device.
|
| |
|