Upon receipt at our post market quality assurance laboratory, this device was thoroughly analyzed.Visual analysis identified that there were heavy burn marks on the connector face, and distorted light was exiting the connector face, indicating an internal fracture.It is likely that the fracture occurred during procedural handling of the fiber during setup or use.This connector fracture can result in an insufficient aiming beam or low laser energy output, up to a complete inability for the fiber to fire.In addition, it is likely that the interaction between the fractured connector and blast shield led to the burn marks observed on the fiber face.The device history record (dhr) confirmed that the device met all material, assembly and performance specifications.There is no objective evidence that the user did not properly handle or use the device according to the instructions for use (ifu).The ifu states: inspect the fiber for kinks, punctures, fractures, or other damage.If the fiber appears damaged, do not use the fiber; return it to the supplier for replacement.Hold the fiber tip to a non-reflective surface, and ensure that a circular red/green spot appears.If the spot is not circular, cleave the fiber.If the spot is weak or not visible, discard the fiber or return it to the supplier for replacement.A risk review was completed and confirmed that the event of fiber damaged/defective was defined in the risk documentation and is documented accordingly.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Based on the information available, a conclusion code of cause traced to component failure was assigned to this investigation.
|