• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LUMENIS LTD SLIMLINE SIS EZ; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LUMENIS LTD SLIMLINE SIS EZ; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number 0644-021-01
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/20/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, this device was thoroughly analyzed.Visual analysis identified that there were heavy burn marks on the connector face, and distorted light was exiting the connector face, indicating an internal fracture.It is likely that the fracture occurred during procedural handling of the fiber during setup or use.This connector fracture can result in an insufficient aiming beam or low laser energy output, up to a complete inability for the fiber to fire.In addition, it is likely that the interaction between the fractured connector and blast shield led to the burn marks observed on the fiber face.The device history record (dhr) confirmed that the device met all material, assembly and performance specifications.There is no objective evidence that the user did not properly handle or use the device according to the instructions for use (ifu).The ifu states: inspect the fiber for kinks, punctures, fractures, or other damage.If the fiber appears damaged, do not use the fiber; return it to the supplier for replacement.Hold the fiber tip to a non-reflective surface, and ensure that a circular red/green spot appears.If the spot is not circular, cleave the fiber.If the spot is weak or not visible, discard the fiber or return it to the supplier for replacement.A risk review was completed and confirmed that the event of fiber damaged/defective was defined in the risk documentation and is documented accordingly.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Based on the information available, a conclusion code of cause traced to component failure was assigned to this investigation.
 
Event or Problem Description
It was reported that the fiber was damaged in its first use after opening the product.The procedure was completed with another device.There were no patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SLIMLINE SIS EZ
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
LUMENIS LTD
yokneam industrial park
hakidma st 6
yokneam 20692 04
IS  2069204
Manufacturer (Section G)
MOLEX LLC
18019 n 25th ave
phoenix AZ 85023
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key25436527
Report Number2124215-2026-30487
Device Sequence Number10322308
Product Code GEX
UDI-Device Identifier07290109145600
UDI-Public(01)07290109145600(17)291111(10)22351224
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K170121
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0644-021-01
Device Catalogue Number0644-021-01
Device Lot Number0022351224
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/17/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/13/2026
Initial Report FDA Received Date06/08/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/12/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-