MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) Style 468 Saline Filled Breast Implant; prosthesis, breast, inflatable, internal, saline

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) Style 468 Saline Filled Breast Implant; prosthesis, breast, inflatable, internal, saline Back to Search Results
Catalog Number 468-560
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/05/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Device has been explanted and replaced.
 
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
 
Event or Problem Description
Healthcare professional reported "deflation and exchange from textured to smooth implant".This record is for the left side.The device remains implanted.
 
Event or Problem Description
Healthcare professional reported "deflation and exchange from textured to smooth implant".Healthcare professional later reported "implant is ruptured and partially calcified".This record is for the left side.The device was explanted.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: h.2., h.3.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: b5, b6, d6b, h6.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: b5, d9, h3, h6.Device evaluation: per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: deflation: observed multiple openings through microscopic inspection 3 openings assessed as surgical damage and 1 opening assessed as fold flaw opening.No further actions are required as it is not related to the manufacturing process.None of the other observations performed during the device analysis creases and wear abrasion are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
 
Event or Problem Description
Healthcare professional reported "deflation and exchange from textured to smooth implant".This record is for the left side.Device has been explanted and replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Style 468 Saline Filled Breast Implant
Common Device Name
prosthesis, breast, inflatable, internal, saline
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS 
Manufacturer (Section G)
Allergan (Costa Rica)
900 parkway global park
zona franca
la aurora de heredia
CS  
Manufacturer Contact
terry ingram
2525 dupont drive
irvine, CA 92612
+1(847)936-6324
MDR Report Key25437485
Report Number9617229-2026-10811
Device Sequence Number21413146
Product Code FWM
UDI-Device Identifier10888628001169
UDI-Public(01)10888628001169(17)191004
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P990074
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/04/2019
Device Catalogue Number468-560
Device Lot Number2826640
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/20/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/12/2026
Supplement Date Received by Manufacturer06/09/2026
08/20/2026
08/31/2026
Initial Report FDA Received Date06/08/2026
Supplement Report FDA Received Date07/02/2026
09/02/2026
09/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/04/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
Patient SexFemale
Patient Weight192 LB
-
-