SMITH & NEPHEW, INC. FAST FIX FLX CRVD INSRTR BNDR CANN SET; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE
|
Back to Search Results |
|
| Catalog Number |
72205324 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/18/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alter the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that, during an arthroscopy, the anchors of a fast-fix flex were deployed successfully.During the suture tightening phase, the surgeon achieved adequate tension, a knot pusher was used in accordance with standard intraoperative technique.During this maneuver, the suture snapped under tension, the device was deemed non-functional for completion of the repair.The pieces were removed with a grasper.The procedure was resumed, after a non-significant delay, using an equivalent s+n back-up.Additional anesthesia was not required.No further complications were reported.
|
| |
|
Search Alerts/Recalls
|
|
|