| Model Number |
SC-8336-70 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Inappropriate Device Stimulation of Tissue (4928)
|
| Date of Event |
05/14/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Block b3: approximated based on the date the manufacturer became aware of the event.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient experienced a rib stimulation.The patient underwent a lead replacement procedure and was doing well postoperatively.The explanted lead will not be returned per hospital policy.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient experienced a rib stimulation.The patient underwent a lead replacement procedure and was doing well postoperatively.The explanted lead will not be returned per hospital policy.
|
| |
|
Additional Manufacturer Narrative
|
|
Block b3: approximated based on the date the manufacturer became aware of the event.Investigation results, media review, device analysis:
the device was not returned for analysis; therefore, a technical analysis could not be performed.Device history record:
it was confirmed this device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Labeling review:
review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Investigation conclusion:
based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
|
| |
|