| Model Number |
AS10 |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/03/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3:product analysis:evaluation could not be performed because of not possible to insert a bur due to detached distal bearings.It was also noted that the color band is faded.G3: aware date used as date of analysis performed date as the event is reported based on evaluation findings.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the attachment cannot attach the tools.It was reported there was no patient impact.Additional information was received stating that detached bearings were found during evaluation.
|
| |
|
Search Alerts/Recalls
|