On-site investigation was performed by our field service engineer.The claimed issue was reproduced during troubleshooting.According to received information in service report, replacement of the maintenance kit including nozzle units solved the issue.The ventilator passed all functional and safety tests and was cleared for clinical use.The nozzle unit is part of the gas module and regulates the inspiratory gas flow to the patient.It consists of a membrane, a mouthpiece and a feather spring.The membrane in the nozzle unit is pressed against a mouthpiece with a feather spring to regulate the gas flow through the gas module.According to the manufacturer¿s specification the complete plastic nozzle unit must be replaced during preventive maintenance.Based on information provided by technician, preventive maintenance had not yet been performed as the unit was only 2 months old at the day of event.The reported issue was confirm in provided device's logs.The defective nozzle unit have not been returned fot the further investigation, despite request.The cause of the claimed issue in this customer product complaint has been determined.The issue was related to defective maintenance kit, including nozzle units.
|