This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: investigation summary the complaint device was discarded by the customer, therefore not returned to depuy synthes and unavailable for a physical evaluation.A manufacturing record evaluation was performed for the finished device lot number (10e5lq), and no non-conformance was identified.Based on the current available information, we cannot determine a root cause for the reported failure.If additional information is received in the future, we will reopen the complaint and perform the investigation as appropriate.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history product number : 222005.Lot number : 10e5lq.There was no non-conformance regarding this lot.Manufacturing date: 11-mar-2026.Expiration date: 28-feb-2029.
|
|
It was reported that during an unknown surgery, the healix advance br 3 suture anchor w/dynacord device, at the knot, the black and white thread is loosened.The blue-white threads and the blue threads held firmly in the anchor, only the one pair of threads had come loose.Another device was used to complete the procedure.Surgery was completed normally.There was no delay in the procedure reported.There were no adverse patient consequences reported.All available information has been disclosed.
|