| Model Number |
694758 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Failure to Capture (1081); Fracture (1260); High impedance (1291); Over-Sensing (1438); Pacing Problem (1439)
|
| Health Effect - Clinical Codes |
Bradycardia (1751); Dizziness (2194); Syncope/Fainting (4411)
|
| Date of Event |
05/17/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Continuation of d10: 429888 lead implanted: (b)(6) 2018; 5086mri45 lead implanted: (b)(6) 2011.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient experienced bradycardia, dizziness, and syncope and was hospitalized.The rv lead exhibited lead integrity alert (lia) warning for non sustained episodes with possible noise.It was also reported that the rv lead exhibited high impedance, no pacing noted in true bipolar or tip to coil pacing attempts due to oversensing and suspected fracture.Device settings were re-programmed with detections off.The rv lead remains in use.No further patient complications have been reported as a result of this event.
|
| |
|