This event was initially deemed not reportable based on our reporting strategy active at the time of the alert date; however, following a recent retrospective review of complaints this event was made reportable out of an abundance of caution.The device history record for lot 30319948 was reviewed and the product was produced according to product specifications.The actual complaint product was returned for evaluation.To examine any possible location of leak or hole, the side port of yellow tubing was flushed through using a 60 ml syringe filled with water.The water immediately began to exit out of the bolus plug.There was not any other location of liquid exit than the bolus plug tip at the distal end.The tip end of yellow tube was pressurized using forceps to observe any potential exit of liquid and still no leakage seen.There was no visible hole anywhere as well after examining the entirety of the yellow tubing.A root cause has not been established.All information reasonably known as of 09 jun 2026 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the reported event but is relaying the information provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21 cfr 803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|