Model Number BL+A217/V806-NIJ
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Unintended Loss of Vascular Access (4867)
Date of Event 04/15/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Customer states that at the end of treatment, upon disconnection of the bloodline, the connection was so tight that a male individual had to apply added pressure to disconnect the blood set from the catheter.The catheter luer broke upon disconnection.2 total incidents.This resulted in hospital admissions for placement of new catheters.No additional information provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIPRO BLOOD TUBING SET W/TP
Common Device Name
BLOODLINE
Manufacturer (Section D)
PT. NIPRO INDONESIA JAYA
kawasan industri surya cipta
jl. surya utama no.23 & 24 kav
karawang,, jawa barat 41363
ID  41363
MDR Report Key25443459
Report Number1056186-2026-00018
Device Sequence Number12216166
Product Code FJK
UDI-Device Identifier00383790011884
UDI-Public00383790011884
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/08/2026
Report Date (Section F) 06/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberBL+A217/V806-NIJ
Device Lot Number25G25L3J
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date04/16/2026
Device Age9 MO
Event Location Outpatient Treatment Facility
Date Report to Manufacturer06/08/2026
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date06/09/2026
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
Date Report Sent to FDA06/08/2026
-
-