| Model Number |
DS760TS |
| Medical Device Problem Codes |
Circuit Failure (1089); Contamination (1120); Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/12/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The reported failure of thermal issue is accommodated in the product risk management file as being linked to hazard with description fire, flame, smoke.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no further indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.The device mentioned in this complaint has exceeded its useful life per the applicable ifu.
|
| |
|
Event or Problem Description
|
|
The bipap auto biflex was returned to a third-party service center for evaluation, and the customer's complaint was not confirmed.During the evaluation, it was noted that the device exhibited water damage and has burnt connector.The root cause isn't confirmed at this time.Additionally, the device was scrapped by the service center after the evaluation was completed.
|
| |
|