|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: a visual and functional assessment was performed on the device.All the pre-repair diagnostic assessment tests passed.During the service evaluation the following failures were identified: no defect found.Conclusion: ¿ failure reported is not confirmed.¿ root cause: no failure found.
|
|
It was reported that the battery handpiece device heated up during use.It was not reported if the device was used in surgery.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2026.All available information has been disclosed.
|