(b)(4).Date sent: 6/09/2026.D4: batch #992c08.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one vasecr35 reload was received with no apparent damage and unfired.The returned reload was loaded into a test device.The returned reload was tested for functionality with a test device in the straight position; however, did not achieved and complete firing sequence.Partially fired 1/10th in the lock-out position with the remaining staples inside the reload pockets.No further functional testing was performed due the condition of the reload.It cannot be confirmed why the reload locked out; however, it is possible that the one-piece sled component within the reload may have been moved out of position during the loading process leading to the inadvertent lockout.As part of ees quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number of 125d04 / 23vasecr35 , and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device batch number 992c08, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|