| Model Number |
6947M55 |
| Medical Device Problem Codes |
High impedance (1291); Under-Sensing (1661)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/03/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Continuation of d10: 5076-45 lead implanted: (b)(6) 2015.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the right ventricular (rv) lead triggered a warning for undefined high pacing impedances and undefined high rv defibrillation coil impedances.The right atrial (ra) lead triggered a warning for undefined high pacing impedances.The left ventricular (lv) lead triggered a warning for undefined high pacing impedances.In addition, the rv lead exhibited undersensing and the ra lead exhibited oversensing.The rv, ra and lv leads remain in use.No patient complications have been reported as a result of this event.
|
| |
|
Search Alerts/Recalls
|