• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. BIPAP A40 PRO, ND; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. BIPAP A40 PRO, ND; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number NDX3100S19
Medical Device Problem Codes Circuit Failure (1089); Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/26/2026
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging the discovery of ventilator inoperative error codes during device evaluation.This issue has been identified under fsn 2023-cc-src-039 and may result in interruptions and/or loss of therapy due to a ventilator inoperative alarm.No patient harm or serious injury was reported in association with this event.The device was not in patient use at the time the issue was identified.The returned bipap a40 pro, nd device was evaluated at a third-party service center.Visual inspection found no evidence of foam degradation.Review of the event log confirmed a ventilator inoperative error code indicating that the difference between the blower pressure sensor reading and the outlet pressure sensor reading was 10 cmh2o or greater for 5 seconds.Based on the evaluation findings, the technician recommended replacement of the device's printed circuit assembly (pca) to address the issue.No repair was performed.The device will be scrapped per customer request.
 
Additional Manufacturer Narrative
The reported failure of ventilator inoperative is accommodated in the product risk management file as being linked to hazard with description loss of function/loss of primary function.A field action was initiated with record id fsn 2023-cc-src-039 and consisted of a field safety notice.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no further indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.Dhr review was performed for the complaint device serial number, (b)(6).Review confirms that the device met the final acceptance criteria and was released for final distribution.The bipap a40 pro (ndx3100s19) is substantially similar to the bipap a40 (1111169) and will be reported in the united states under bipap a40, 501k number: k121623.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPAP A40 PRO, ND
Common Device Name
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key25447959
Report Number2518422-2026-020350
Device Sequence Number12222526
Product Code MNT
UDI-Device Identifier00606959059092
UDI-Public(01)00606959059092
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K121623
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 06/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberNDX3100S19
Device Catalogue NumberNDX3100S19
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/26/2026
Initial Report FDA Received Date06/09/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/12/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1813-2024
Patient Sequence Number1
Patient SexUnknown
-
-