| Model Number |
693565 |
| Medical Device Problem Codes |
Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574)
|
| Health Effect - Clinical Codes |
Muscle Weakness/Atrophy (1967); Pain (1994); Dizziness (2194); Shock from Patient Lead(s) (3162)
|
| Date of Event |
05/18/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Continuation of d10: 5524m-53 lead, implanted: (b)(6) 2001.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the right ventricular (rv) lead exhibited t-wave oversensing (twos) post-ventricular pace, leading to possible inappropriate therapy.The patient was asleep at the time of the event, but upon waking up they reported feeling like they had hit their head and feeling "a little woozy, but otherwise fine." the lead remains in use.No further patient complications have been reported as a result of this event.
|
| |
|
Event or Problem Description
|
|
It was further reported that the patient reported "feeling weak, then fell to their knees and then got a shock".
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: h6 (annex a) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|