| Model Number |
414005 |
| Medical Device Problem Codes |
Break (1069); Ambient Noise Problem (2877); Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/09/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Combination product: yes.Biotronik submits this report to comply with fda regulations 21 cfr parts 4 and 803.Biotronik has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by biotronik, or its employees that the device, biotronik, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Biotronik will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
This rv lead was explanted due to noise, high pacing impedance, increasing shock impedance and is-1 pin disconnected from ring.No adverse patient events were reported.Should additional information be received, this file will be updated.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes.The lead is currently not available for analysis.No conclusion can be drawn based on available information at the moment.The file is closed.The investigation will be re-opened should additional data become available.
|
| |
|