MDT POWERED SURGICAL SOLUTIONS MIDAS REX; MOTOR, DRILL, PNEUMATIC Back to Search Results
Model Number AF02
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/21/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3: product analysis: evaluation determined that internal tube bearing was sticking or stuck.The likely cause of failure couldn't be able to determined.It was also noted that tool cannot be inserted and cannot perform temperature test.This regulatory report is being submitted due to retrospective review through capa 672570.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the attachment bearing has fallen off.It was reported there was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDAS REX
Common Device Name
MOTOR, DRILL, PNEUMATIC
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer (Section G)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
7635265677
MDR Report Key25452890
Report Number1625507-2026-02320
Device Sequence Number13349151
Product Code HBB
UDI-Device Identifier00763000084738
UDI-Public(01)00763000084738(11)190703
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K170312
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAF02
Device Catalogue NumberAF02
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/03/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/21/2025
Initial Report FDA Received Date06/10/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/03/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-